GE NBP Pump Assembly 801491-002
NBP NIBP Pump Assembly 801491-002 for GE Dash3000 Dash4000 Dash5000 Patient Monitor
► Brief Description
NBP NIBP Pump Assembly 801491-002 for GE Dash3000 Dash4000 Dash5000 Patient Monitor
Replace NBP pump assembly:
1. Remove the DAS assembly.
2. Remove the four screws and remove the NBP assembly from the DAS assembly cover.
3. Disconnect the tubing going to the internal solenoid valve of the NBP assembly.
4. Using a small flat-blade screwdriver, remove the connector from the NBP assembly to the DAS assembly.
5. Connect the NBP component to the DAS component. Don't forget: the four mounting screws, the tube connection, and the flex connector.
6. Reassemble the patient monitor in reverse order.
► Specification
Brand |
GE |
Product Category |
Monitoring |
Product Type |
Spare Parts |
Product Name |
NBP / NIBP Pump Assembly |
Part Number |
801491-002 |
Machine Application |
Dash3000 Dash4000 Dash5000 |
Condition |
Used - Good |
Warranty |
90 Days |
Delivery Date |
3 - 5 Working Days |
Package |
Carton |
In Stock |
Yes |
Price |
Accept Negotiation |
MOQ |
One Piece |
Material |
Plastic, Metal, etc. |
Payment Method |
Western Union, Bank T/T, Paypal |
||
Transpotation |
DHL, FedEx, UPS, TNT, EMS, By Air, etc. |
► Detailed Photos
Thanks for your time, if you are interested in NBP NIBP Pump Assembly 801491-002 for GE Dash3000 Dash4000 Dash5000 Patient Monitor, or have any suggestions for us, please feel free to contact us. We will reply you within 12 hours, wish you a happy life.
Tags: NBP NIBP Pump Assembly 801491-002 for GE Dash3000 Dash4000 Dash5000 Patient Monitor, NBP Pump Assembly 801491-002, 801491-002,
Copyright statement:If not specified, the articles on this site are all PMR Med originals. Please indicate the source and attach the link to the NBP NIBP Pump Assembly 801491-002 for GE Dash3000 Dash4000 Dash5000 Patient Monitor article for reprinting.
Previous:CHILWEE Battery 6-FM-4.5 12V, 4.5Ah for Philips Goldway UT4000F PRO Patient Monitor
Next:Edan iM8 Patient Monitor Right Side Cover 21.51.410121, 01.51.30273 / Recorder’s Paper Inlet Cover
Send your inquiry directly to us
Contact Person:Katie Lee
Tel:8618879071075
Fax:86-188-790-71075